Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med PM03: CAL ADJ /5500,34401 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 2 NUMBER OF LINES: 59 CONFIGURATION: HP 34401A CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Astro-Med PM03 Calibration Procedure Rev "A" # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 20 min. - REQUIRED 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <75% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {FULL SCALE AMPLIFIER CALIBRATION (RMS MODE)} 1.008 DISP Press SYSTEM DEFAULT on the UUT. 1.009 DISP Connect the 34401A to the UUT as follows: 1.009 DISP [32] 34401A Input HI to TP1 1.009 DISP [32] 34401A Input Lo to GROUND 1.009 DISP 1.009 DISP Connect the 5500A to PM03 INPUT. 1.010 DISP Set the UUT to the 10V range. 1.011 MESS Adjust R98 for 0mV +/- 5mV. 1.012 34401 0mV S 2W 1.013 DISP Set the UUT to RMS mode. 1.014 5500 10V 60H SI S 2W 1.015 EVAL Is display 17 division from center +/-1/2 div/? 2.001 END